Suggested dimensions of the table top are:
30 in. deep
48 in. wide
36 in. high
Figure 2-1: A typical dose preparation layout.
Frequently, dose calibrators are located in areas called hot
labs in which there is a wet sink with plumbing. Do not
locate your dose calibrator on a table or other counter
which can become wet or is subject to splashing or spray-
ing.
Avoid areas which have drafts caused by heating or air con-
ditioners, or are in direct sunlight. Electronics work best
when constant temperature is maintained.
The temperature range is from 0 to 40° C, and the relative
humidity range from 0 to 95%.
Choose a location for your dose calibrator which is not fre-
quently used by other personnel one where the table on
which it lies will not be bumped.
In general, make sure your work area is clean, dry, and dust
free.
ATOMLAB Dose Calibrator Operation Manual Chapter 2
Biodex Medical Systems 2-2 Release 2.0
Physical Location
and Environment