ATOMLAB Dose Calibrator Operation Manual Packing Slip
PACKING SLIP
The Atomlab 100 Dose Calibrator (#086-250 or #086-258
with RS-232 port) and its respective components comes
carefully packed in two cartons inside a large, single ship-
ping box. The cartons, each marked with an appropriate
serial number, are labeled:
DISPLAY UNIT
DETECTOR UNIT
Prior to removing any of the cartons from the shipping
box, visually inspect the box and the inside packing for dam-
age. If any damage is noted, do not unpack the contents.
Notify the carrier immediately so that a claim can be made if
necessary, and contact your Biodex Medical Systems Sales
Representative. If there is no visible damage, remove the
cartons and place them on a table or to the side. We suggest
removing the DISPLAY UNIT carton first.
Before deciding on a location to set up your Atomlab
for operation, be sure to read Chapter Two on installation.
This chapter notes operating and environmental require-
ments, and offers several tips on choosing a location for
your dose calibrator.
In addition to the Display Unit itself, this carton contains the
Atomlab's Operation Manual, all the necessary connection
and power wires needed to run the system, and a Moly
Assay Kit. A Report of Calibration can be found in the
Instruction Manual's front, inside pocket.
The carton labeled DETECTOR UNIT is very
heavy, ~32 pounds. Severe damage to the unit may result if
this carton is dropped. Remove the detector from the carton
and carefully place it on the table or other counter where you
will set up and use your Atomlab unit. Make certain the table
top will support a minimum of 300 pounds. We suggest that
you place the carton on its side so that you can slide the
Detector Unit out of the carton onto its base. Inserted into the
Chamber Well should be a Chamber Well Liner. A Plastic
Sample Holder is packed with the Display Unit.
Shipping Cartons:
The DISPLAY
UNIT Carton:
The DETECTOR
UNIT Carton:
Biodex Medical Systems I Release 2.0