DETECTOR TYPE
Well-type pressurized ionization chamber, with Argon filled
gas. Replaceable plastic liner included. Dimensions are:
Well opening: 2.5 in. (6.4 cm)
Well depth: 10.25 in. (26 cm)
.25 in. (6 mm) lead surrounding ionization chamber, with top
well opening.
300 volt electronic power supply.
ENVIRONMENTAL OPERATING CONDITIONS
Temperature: 0 to 40°C
Humidity: 0 to 95% rh, non-condensing
For optimum performance, the Atomlab 100 should be oper-
ated in a normal laboratory environment where the tempera-
ture and humidity are maintained for normal human comfort
and the ambient radiation level is low and constant.
PHYSICAL DATA
Display Unit Detector Unit
Size: 3.75 x 12 x 14.3" 7.5 x 7.5 x 16.25"
(9.5 x 30.5 x 36.3 cm) (19 x 19 x 41.3 cm)
Weight: 6 lb (2.72 kg) 35 lb (15.9 kg)
POWER
100 to 120 VAC 1/2 A, or 200 to 240
VAC 1/4 A, selectable with fuse card.
50/60 Hz
Detachable line cord. Built-in EMI filter and transient sup-
pression.
ATOMLAB Dose Calibrator Operation Manual Appendix B
Biodex Medical Systems B-3 Release 2.0