APPENDIX B
Atomlab 100 Dose Calibrator Specifications
OVERVIEW
This appendix provides the physical and operating specifi-
cations of your Atomlab 100 Dose Calibrator.
MEASUREMENT RANGE
For Tc-99m, 0.01 µCi to 9999 mCi, auto-ranging.
DISPLAY RANGES
Units are user selectable:
Ci Bq
00.01-19.99 µCi .001-1.999 MBq
20.0-199.9 µCi 2.00-19.99 MBq
200.-1999. µCi 20.0-199.9 MBq
2.00-19.99 mCi .200-1.999 GBq
20.0-199.9 mCi 2.00-19.99 GBq
200.0-9999. mCi 20.0-399.9 GBq
RESPONSE TIME
Auto selection:
one second above 2 mCi;
three seconds between 200 µCi and 2 mCi;
three to 30 seconds below 200 µCi.
ACCURACY
Overall accuracy of activity determination for specific front
panel isotopes using the factory set Calibration Values is
within 3% or 0.3 µCi, whichever is greater.
: Intercomparison of calibrated sources and the
Atomlab 100 requires combining the uncertainties of both
the Atomlab 100 and the calibrated source to determine
expected agreement.
ATOMLAB Dose Calibrator Operation Manual Appendix B
Biodex Medical Systems B-1 Release 2.0