Within 1% or 0.2 µCi, whichever is greater.
Within 1% or 0.2 µCi, whichever is greater.
Within 1% or 0.2 µCi, whichever is greater.
Short term (24 hours): within 0.3% above one mCi,
exclusive of background.
Long term (one year): within 1%.
NOISE
Display Fluctuations Activity Range
< 1 Digit > 199.9 µCi
< ±2 > 19.99 µCi
< ±4 0.01 to 19.99 µCi
ENERGY RANGE
25 keV to 3 MeV
ISOTOPE SELECTION
Ten preset Keys: Co-57, Cs-137, I-131, In-111, Ga-
67, Xe-133, I-123, Tl-201, Mo-99, and Tc-99m.
Two user programmable keys: 1 and 2.
OTHER key for setting on-the-spot Calibration Values for
non-standard isotopes.
CALIBRATION VALUE DISPLAY
Four digit Calibration Value Display with
Increment/Decrement keys to change value. Range is from
000.1 to 999.
ATOMLAB Dose Calibrator Operation Manual Appendix B
Biodex Medical Systems B-2 Release 2.0
Detector
Linearity
Electrometer
Linearity
Electrometer
Accuracy
Stability