The second group of commands is included to aid developers
of third party software. The characters "1,2,3" will not be inter-
cepted or interrupted by off-the-shelf serial port software pack-
ages, whereas the control characters very likely will be.
Do not attempt to communicate with the calibrator during its
power-up sequence. The response is undefined. All PC port
commands may be ignored after this attempt.
During normal operation, the above six characters are recog-
nized, all others are ignored.
During a self-test sequence, i.e., the TEST key was pressed, all
characters are ignored, including the six PC port commands.
The PC port operates at 4800 baud, no parity, 8 data bits and
1 stop bit.
The start of a record character is a tilde (0x7E ~ ~) (Hex Ascii
Key). The end of a record character is a line feed (0x0A LF
^J). A record is made up of four fields in the case of the Dose
Calibrators and five fields in the case of PET calibrators. The
field separator is one or more spaces (0x20 SP Spacebar).
The first field starts with the first character after the title and
continues until a space is encountered. It may be from two till
six characters in length. There are thirteen isotope names.
Typical sample records are shown below (version 5 firmware):
Atomlab 100
~Tc-99m 31.5 2.54 µCi
~Mo-99 179. 25.6 mCi
~Tl-201 18.0 14.0 MBq
~I-123 14.4 19.5 GBq
~Xe-133 19.6 1.57 µCi
~Ga-67 28.0 2.24 µCi
~In-111 13.5 -1.08 µCi
~I-131 22.6 1.80 µCi
~Cs-137 17.3 1.37 µCi
~Co-57 27.0 2.15 µCi
~#2 278. 22.3 µCi
~#1 5.0 0.40 µCi
~OTHER 7.7 -0.61 µCi
ATOMLAB Dose Calibrator Operation Manual Appendix G
Biodex Medical Systems G-2 Release 2.0
Records
Dose Calibrator
Records