When you press one of the User Defined Isotope
Selection keys (the OTHER, 1 and 2 keys), the Calibration
Value for the selected key will always appear on the
Electronic Thumbwheel Display. The only time the
Calibration Value for the other Isotope Selection keys will
appear on the Electronic Thumbwheel is when the Atomlab
100 is in Setup mode.
1) Press the SETUP key. The SETUP LED comes ON. The
OTHER key LED begins to flash indicating you may
select the OTHER display units if desired.
2) Press the key labeled 1 (or 2). The 1 (or 2) LED comes
ON.
3) Select the Calibration Value from Appendix E
corresponding to the radioisotope you intend to
measure.
4) Use the
⇑⇓
keys to set the Calibration Value on the
Electronic Thumbwheel Display. The ENTER LED starts
flashing.
5) Press the ENTER key. The 1 (or 2) key is now
programmed and will retain the Calibration Value you set
it to, even if you turn OFF your Atomlab 100.
6) Press the SETUP key. The Atomlab 100 is now out of
Setup mode.
Should you need to return the standard Isotope Selection
keys (labeled Co-57, Cs-137, ... , Tc–99m) to their original
preset Calibration Values, follow the above steps, but with
the following changes:
For step 2) press the standard Isotope Selection key in
question.
For step 3) look up the preset calibration dial value in
Appendix D.
ATOMLAB Dose Calibrator Operation Manual Chapter 5
Biodex Medical Systems 5-8 Release 2.0
The Standard
Keys