Two round Increment/Decrement keys below display. When
an Increment/Decrement key is depressed momentarily, the
Display changes accordingly by one digit.
If the key is held down, the Display changes slowly at first
(1 digit per 1/4 second), then more quickly after two-and-
one-half seconds (ten digits per 1/4 second), after which the
display will change full range in about seven seconds.
Displays Calibration Value when OTHER, 1 or 2, has
been selected;
Error Number when ER. is lit.
Calibration Value of pre-selected isotope when in Setup
mode (000.1-99.).
The ENTER key (to the right of the Electronic Thumbwheel
Display) programs the Calibration Value to correspond with
the selected isotope key.
The SETUP key allows the selection of activity display units
and the calibration of all of the isotope Selection keys. To
enter Setup mode, press the key once. Pressing the key a
second time takes the Atomlab 100 out of Setup mode.
The TEST key enables the Dose Calibrator’s Self-Test pro-
gram, which includes a test of the electrometer sensitivity,
chamber bias supply, software checksum, and the display.
The BKGND key enables the background radiation measure-
ment for storage and automatic correction.
100 to 120 VAC 1/2 A, or 200 to 240 VAC 1/4 A, selectable
with Fuse Holder (See Appendix A).
Electrical Equipment, Laboratory Use;
Part 1, General Requirements for Safety conforms to
UL STD 3101-1, conforms to EN 6010-1,
EN 60601-1-1
ATOMLAB Dose Calibrator Operation Manual Appendix B
Biodex Medical Systems B-5 Release 2.0
Electronic
Thumbwheel
Display
Control Function
Keys
Line Voltage
ETL Listed