E30 Digital Electrocardiograph User’s Manual
2-3
voltage and frequency ratings of the power cord are the same as those
indicated on the cardiograph’s label or in this manual.
Always install or carry the cardiograph properly to avoid damage
caused by drop, impact, strong vibration or other mechanical force.
Note:
Put the cardiograph in a location where you can easily see the screen
and access the operating controls.
Keep this manual in the vicinity of the cardiograph so that it can be
obtained conveniently when needed.
The software was developed in compliance with IEC 60601-1-4. The
possibility of hazards arising from software errors is minimized.
This manual describes all features and options. Your cardiograph may
not have all of them.
2.2 General Safety
Warning: The cardiograph is neither a therapeutic instrument nor a
device that can be used at home.
1.Safety precautions for installation
z
Connect the power cord to a properly earthing socket. Avoid putting the
socket used for it in the same loop of such devices as the air conditioners,
which regularly switch between on and off.
z
Avoid putting the cardiograph in the locations where it easily shakes or
wobbles.
z
Enough space shall be left around the cardiograph so as to guarantee
normal ventilation.
z
Make sure the ambient temperature and humidity are stable and avoid the
occurrence of condensation in the operation process of the cardiograph.
Warning: Never install the cardiograph in an environment where
flammable anesthetic gas is present.
0
))