E30 Digital Electrocardiograph User’s Manual
22
E.4.2 Disclosed changes of measurments caused by NOISE on ECGs
Global
measurement
Type of added
NOISE
Disclosed differences
Remark
Mean (ms)
Standard deviation
MA_ MA_
P-duration
High 5.8 8.4
AC 0.2 7.2 50Hz
AC 6.0 12.0 60Hz
Baseline 5.6 8.4
PR-interval
High 2.4 8.2
AC -5.6 7.7 50Hz
AC 1.6 6.7 60Hz
Baseline 0.0 9.1
QRS-duration
High 3.2 7.9
AC 2.2 6.4 50Hz
AC 2.0 10.2 60Hz
Baseline 2.2 9.5
QT-interval
High 4.2 7.6
AC 7.4 7.9 50Hz
AC 5.0 6.9 60Hz
Baseline 3.6 8.7