E30 Digital Electrocardiograph User’s Manual
2-7
close patient observation or alternative monitoring devices should be used
until electrocardiograph function is restored.
If the electrocardiograph does not automatically resume operation within
60s, restart the electrocardiograph using the power switch. Once monitoring is
restored, you should verify correct monitoring state and alarm function.
Intended for use in conjunction with other medical devices
The electrocardiograph can be used together with high-frequency
electrotomes and defibrillators. Patient’s safety must be ensured during the
process.
Prompt
The machine can give prompts of abnormal status arising from excessive
polarization voltage of ECG electrode.
For any question, contact us or native agent.
2.4 Safe Operation Conditions
Methods of sterilization or
disinfection recommended by
the manufacturer
Sterilization: not applicable
Disinfection: Refer to Clean and Maintenance
Chapter
Electromagnetic interference
No mobile telephone nearby
Electrosurgical interference
damage
No damage
Diathermy instruments
influence
Displayed values and prints may be disturbed or
erroneous during diathermy
Defibrillation shocks
The electrocardiograph specifications fulfill the
requirements of IEC 60601-2-25.