E30 Digital Electrocardiograph User’s Manual
4-7
Operating refers to the following steps:
1. Before recording, enter the patients’ info.
2. Press “ ” on the panel to switch the working mode into Manual mode.
3. After select 【Work Mode】 in main interface, you can select 【Manual
style】 according to actual needs, then press 【OK】 to complete the setup.
Or you can just press 【Man1】 to select the proper report format below
the monitor interface directly.
4. Press “ ” on the panel to start recording.
5. To stop recording in the course of recording, press “ ” on the
panel.
4.5.3 Rhythm Mode
In Rhythm mode, users may select lead according to actual needs, and
record the rhythm waveform of single lead within 60s or rhythm waveforms of
3 leads within 20s each.
Operating refers to the following steps:
1. Before recording, enter the patients’ info.
2. Press “ ” on the panel to switch the working mode into Rhythm mode.
3. After select 【Work Mode】 in main interface, you can select 【Rhythm
style】 and 【Rhy-lead】 according to actual needs, then press 【OK】
to complete the setup. Or you can just press 【Rhy1】 to select the proper
report format below the monitor interface directly.
4. Press “ ” on the panel, the prompt info area will display
“Sampling…”, and timing the sample time. It will start to draw when the
time reaches 60s or 20s.
5. The machine will stop automatically after a complete rhythm diagram is
recorded.
6. To stop recording in the course of recording, press “ ” on the
panel.
Caution: Please check whether there is any error message shown on
the screen before start recording. If electrodes fall off displayed,
check whether the corresponding electrode is well connected with
patient or the polarized voltage overtops. If unconnected electrodes
displayed, check whether RA electrode is well connected with