E30 Digital Electrocardiograph User’s Manual
1
Appendix A Product Specifications
A.1 Safety Specifications
According to the MDD 93/42/EEC, the electrocardiograph is Type a equipment.
Classified according to the IEC60601-1 is as follows:
Classification of
protection against
electric shock
Class I, internally and externally powered equipment.
Degree of protection
against electric shock
Type CF applied part, and is defibrillator-proof.
Degree of protection
against ingress of
liquid
Ordinary equipment
Degree of protection
against hazards of
explosion
Equipment is not suitable for use in the presence of
flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Mode of operation
Continuous
Electromagnetism
compatible
Group I Class B
A.2 Environmental Specifications
Operating
Temperature
5℃~40
Relative humidity
85% (non condensing)
Atmospheric
pressure
700hPa1060hPa
Transportation and storage
Temperature
-20℃~+55
Relative humidity
93% (non condensing)
Atmospheric
pressure
500hPa1060 hPa