E30 Digital Electrocardiograph User’s Manual
7-4
Caution: Disinfection possibly causes damage to Electrocardiograph
to some extent. It is suggested that disinfection be conducted only
when your hospital maintenance plan deems it as necessary. The
equipment shall be cleaned before disinfection.
Note: Electric schematic diagram and list of components are provided
to the qualified maintenance station or personnel certified by the
manufacturer only.
7.4 Cleaning and Maintenance of Recorder
To prevent stain on surface of thermosensitive printing head due to
excessive long period of use of printer and clarity of recording is adversely
affected, the users shall regularly (at least once every month) clean the surface
of recording head:
Open the box cover of recorder;
Take out the remaining chart paper;
Gently wipe the surface of recording head by use of clean soft cloth dipped
with small quantity of diluted alcohol;
Air dry the recorder at a cool and ventilated place;
Place the chart paper properly, and close the box cover of recorder.
)