E30 Digital Electrocardiograph User’s Manual
7-3
e) Air dries the cable at a shady and cool place.
Caution:
It is necessary to regularly check the cable and lead wire of the
patient to ensure their completeness and good conduction;
Damaged or aging cable or wire shall be replaced timely;
In the course of use, tidy up the lead wire as much as possible, and do
not pull it forcibly.
2. Chest electrode and limb electrode
a) Please clean chest electrode and limb electrode before disinfection.
b) Wipe off the conductive paste on surface of electrode with soft cloth;
c) Take apart the electrode plate and clamp of limb electrode as well as rubber
ball and metal cup of chest electrode;
d) Put electrode into clean warm water (not higher than 35℃) and clean it to
ensure no residue of conductive paste;
e) Air dries the electrode at a shady and cool place.
Caution:
Electrode shall be timely cleaned after use;
Rubber ball of chest electrode shall be prevented from direct sunlight;
otherwise, aging will be caused;
Electrode with eroded surface shall be timely replaced with new
electrode.
Caution:
Do not disinfect cable and lead wire with high voltage, radial or steam.
Do not dip cable or lead wire directly in liquid.
To prevent long-term damage to cable, it is suggested that the product be
disinfected only when your hospital regulation deems it as necessary.
Do not clean or reuse disposable electrode.
Warning: Do not use EtO, phenyl, amido or iodo for disinfection of
the machine.