E30 Digital Electrocardiograph User’s Manual
2-6
Do not use this equipment in the presence of flammable anesthetics, vapors
or liquids.
Leakage current test
When interfacing with other equipment, a test for leakage current must be
performed by qualified biomedical engineering personnel before using with
patients.
Battery
The device is equipped with a battery. The battery discharges even when the
device is not in use. Store the device with a fully charged battery and take out
the battery, so that the service life of the battery will not be shortened.
Disposal of accessories and device
Disposable accessories are intended for single use only. They should not be
reused as performance could degrade or contamination could occur.
When operating with defibrillator at the same time, cables with
defibrillator-proof must be adopted in order to avoid any unwanted damage.
The service life of this electrocardiograph is 5 years. At the end of its
service life, the electrocardiograph, as well as its accessories, must be
disposed of in compliance with the guidelines regulating the disposal of such
products. If you have questions concerning disposal of products, please
contact manufacturer or its representatives.
EMC
Magnetic and electrical fields are capable of interfering with the proper
performance of the device. For this reason, make sure that all external devices
operated in the vicinity of the electrocardiograph comply with the relevant
EMC requirements. X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation. Also,
keep mobile phones or other telecommunication equipment away from the
electrocardiograph.
Instruction for use
For continuous safe use of the electrocardiograph, it is necessary that listed
instructions were followed. However, instructions listed in this manual in no
way can supersede established medical practices concerning patient care.
Loss of data
Should the electrocardiograph at any time temporarily lose patient data,