Rev A N-7
8. Change Reference Ranges / Thresholds for Tests:
a. Select SET PARAMETERS using the keys and Press the key.
b. Select
RANGES using the keys and Press the key.
c. Select the test panel type using the keys.
d. Select the analyte and level using keys. Type in the new value for
the
HIGH or LOW cutoff as appropriate.
e. Press the key to save changes.
9. Change QC Parameters:
a. Select SET PARAMETERS using the keys and Press the key.
b. Select
QC PARAMETERS using the keys and Press the key.
c. Select the desired setting using the keys and change
the value to the right of the setting by using the keys.
Setting Default Options New Setting
QC DEVICE FREQ. Daily None, 8HR, Daily,
Weekly, Monthly
QC SAMPLE FREQ. Monthly None, 8HR, Daily,
Weekly, Monthly
NUMBER OF CONTROLS 2 1 or 2
d.
Press the
key to save changes to QC Parameters.
For more details
about Ranges Settings
s
ee page N-24.
For more details about
QC Parameters Settings
see page N-27.
N
ote: The lowest
value of the
Range or Threshold
may be disabled for
some test panels. Refer
to the appropriate
Product Insert for
specific information.