Rev AO-4
MAIN MENU
RUN TEST
1
0. The Triage MeterPro pulls in the QC Device and scans it. The test
d
evice may partially move in and out of the meter several times.
11. When the test is complete, the meter will beep, eject the device and display the
results on the meter’s screen.
12. Press the key to make a printed copy of the results.
IM
PORTANT INSTRUCTIONS FOR ALL TESTS
:
a. Gently insert the test device into the Triage MeterPro until you feel
the
device catch on the pin. You will hear an audible ‘click’ that tells you
the test device has been inserted properly.
b. Press the key to start the test.
c. The Triage MeterPro pulls in the test device and scans it.
d. When the test is complete, the meter will beep and display the results
on the meter’s screen.
e. Press the key to make a printed copy of the results.
IMPORTANT INSTRUCTIONS FOR QC DEVICE TESTS:
Dust, lint, fibers and other small particles may interfere with the QC Device.
Keep the QC Device free of contaminants.
The QC Device is light-sensitive and should be stor
ed in its black
opaque case when not in use.
If the QC Device tests fail, wipe the QC Device clean with a lint
fr
ee cloth to remove any oils, dust, fibers, or fingerprints. Do not apply
any liquid to the QC Device. After cleaning the device, r
epeat
the QC Device test.
If the QC Device test fails after you have cleaned the device, call
Biosite Incorporated Technical Ser
vices at 1-888-BIOSITE.
Note: the QC Device does not expir
e.
After the test
device has been
inserted, do not push
the device in further or
attempt to pull it out.
The device may be
ejected by returning to
the Main Menu and
pressing the
EXIT key.