Predefined Protocols Listing| 41
Predefined Protocols Listing
LHCUsers: Folders named 4Buffers and 1Buffer contain customized protocols for
models with and without Buffer Switching, respectively.
Maintenance Protocols
Daily
Maintenance
Description
W-DAY_
RINSE
Simple one-step protocol to fully flush the system with water or
reagent to keep the manifold tubes clog-free. Defined for use with
Buffer A; 500 mL total volume.
W-
OVERNIGHT_
LOOP
Protocol designed to keep the manifold in a wetted condition
overnight or for a long downtime period; manifold tubes are
submerged in fluid for 4-hour intervals between primes in this virtually
endless loop. Defined to use Buffer A.
W-RINSE_
AND_SOAK
Identical to W-DAY_RINSE with one addition, the manifold tubes are
submerged and soaked for 5 minutes in the fluid.
Periodic Maintenance
W-
Decontaminate
Two/three stage protocol to decontaminate the washer, first
flushing lines with disinfectant and then rinsing them with
deionized or distilled water. Buffer switching models flush and
then rinse all four lines. Manifold tubes are submerged and
soaked for 20 minutes during each pass. Prompts guide the user
through the process.
W-LONG_
SHUTDOWN
Helps implement the recommended procedure for preparing the
instrument for storage. This protocol includes prompts for running
disinfectant from Buffer A, then water from Buffer B, and lastly,
air through the system - remove bottle from Buffer C valve.
W-PRIME_
250/300
Simple prime routine; defined for Buffer valve A only. For LHC
only.
PRIME_ALL_
BUFFRS
Consecutively primes each of the Buffer Switching valves
beginning with D. Designed for use in the annual instrument
verification test of the Buffer Switching module.