CE Mark| 89
Caution: High Flow Pump Installation. DO NOT plug the High Flow
vacuum pump cable into a wall outlet! Use the adapter provided with the
pump to connect the pump to the accessory outlet on the back of the
washer. See the Installation instructions.
Caution: WasteSensorPort on 405 TS.(For customers who have
purchased the BioStack Microplate Stacker.) Although the waste sensor
port on the back of the 405 TS is the same type as the 24-VDC power
connector on the back of the BioStack, if an external 24-VDC power
supply is plugged into the 405 TS’s port, it will permanently damage
internal components.
Caution: Electromagnetic Environment. Per IEC 61326-2-6 it is the
user’s responsibility to ensure that a compatible electromagnetic
environment for this instrument is provided and maintained in order that
the device will perform as intended.
Caution: Electromagnetic Compatibility. Do not use this device in
close proximity to sources of strong electromagnetic radiation (e.g.,
unshielded intentional RF sources), because these may interfere with the
proper operation.
CE Mark
Based on the testing described below and information
contained herein, this instrument bears the CE mark.
l Note: See the Declaration of Conformity for specific information.
Directive 2004/108/EC: Electromagnetic Compatibility
Emissions—Class A
The system has been type-tested by an independent, accredited testing laboratory
and found to meet the requirements of EN 61326-1: Class A for Radiated Emissions
and Line Conducted Emissions.
Verification of compliance was conducted to the limits and methods of EN 55011
(CISPR 11) Class A. In a domestic environment it may cause radio interference, in
which case, you may need to mitigate the interference.
Immunity
The system has been type-tested by an independent, accredited testing laboratory
and found to meet the requirements of EN 61326-1 and EN 61326-2-6 for Immunity.
BioTek Instruments, Inc.