42 | 405 TS Getting Started Guide
QC (Quality Control) Protocols
Manifold-Specific
QC_96_DISP_TEST Dispense precision test protocol for 96-tube manifold.
QC_96_EVAC_
TEST
Evacuation efficiency test protocol for 96-tube manifold.
QC_192_DISP_
TEST
Dispense precision test protocol for 192-tube manifold.
QC_192_EVAC_
TEST
Evacuation efficiency test protocol for 192-tube manifold.
QC_96_VAC30_
TEST
Vacuum filtration evacuation efficiency test for 96-well filter
plates and also recommended for use in maintenance
procedures.
QC_384_VAC10_
TEST
Vacuum filtration evacuation efficiency test for 384-well
filter plates and also recommended for use in maintenance
procedures.
Predefined Sample Protocols
The "Sample" protocols are provided to facilitate learning. Some samples are model
specific.
LHC users: First open the folder that matches your washer model, e.g. 405 Select.
You may need to customize the protocols (as described on page 39) to match your
instrument's settings.
Cell Wash
W-
CELLWASH_
96
Cell wash protocol designed to minimize cell layer disturbance in
96-well plates. Aspirate height increased to 50 steps (approx.
6.35 mm above plate carrier) resulting in increased residual,
approximately 100 µL.
W-
CELLWASH_
384
Cell wash protocol designed to minimize disturbance to cells in
384-well plates. Aspirate height increased to 50 steps (approx.
6.35 mm above plate carrier) will cause increased residual
volume but help to preserve the cell monolayer.