68 | 405 TS Getting Started Guide
LHC
Touch screen
1. Select Tools>
Instrument Utilities.
2. For Plate Carrier
Selection, select Vacuum
Filtration.
1. Press Instrument >
Options
2. Set the Plate Carrier to
Vacuum Filtration.
Note: Make sure Vacuum Filtration is "enabled" to select the special plate carrier:
Instrument>Options>Advanced Settings.
Important: Always set the carrier selection to match the installed hardware, regardless
of the type of plate processing you are doing. You can perform a regular wash (non-filter
plates) using the vacuum filtration carrier.
Changing the Instrument's Settings
About the Onboard Settings
The instrument's onboard settings dictate its behavior. Certain settings, like Plate
Clearance, can improve performance. Other settings are critical to its performance.
For example, the manifold setting must match the installed manifold type.
LHC users: do not confuse the LHC's Target Instrument Settings assigned to each
protocol with the onboard settings! Review the Help topic: LHC Protocols Explained to learn
the distinctions.
To change the onboard settings:
LHC Touch screen
1. Select Tools>Instrument
Utilities
2. Choose the applicable tab and
modify settings as desired.
3. Click Send after changing a
setting to update the
instrument.
1. Press
Instrument>Options
for washer settings.
2. Explore the other tabs
for more settings.
l When using the LHC to change instrument settings, you may need to
reboot the 405 TS before controlling it with the touch screen.