H-80 Series User Manual V1.0
33 / 46
When the device works normally, if the
inlet port is blocked, the device will give
the inlet port blockage alarm; the screen
will display ―Inlet Port Blocked!!‖
If the device detects that the output
voltage supplied by medical power is lower
than 22V, an audible alert will sound and
the screen will display ―Low Input
Voltage!!‖.
Heated Humidifier
Failure!!
The device starts the heated humidifying
function, but the temp does not rise after
5min, the device gives an alarm of
humidifier failure, and ―Humidifier
Failure!!‖ is displayed on the display
screen.
Give an alarm within
10min
The device detects that the output voltage
is lower than 22V and sends an alarm of
low voltage, and the screen will display
―Low voltage !!‖.
Immediately give an
alarm without delay
Can’t Reach Target
FiO
2
!!
When the device detects that the set
target FiO
2
can't be reached, it gives an
alarm that the target FiO
2
can’t be
reached, and the screen will display
―Can't Reach Target FiO
2
!!‖.
(applicable to H-80A and H-80AS)
Can’t Reach Target
Temp!!
When the device detects that the set
target temp can’t be reached over 2 ℃, it
gives an alarm then the screen will display
―Can’t Reach Target Temp!!‖.
Start to judge 30s
after starting heating,
and continue for
15min. If the judge
temperature is still
lower than the set
target temperature
over 2 °C, the device
give an alarm
Can't Reach Target
Flow!!
When the device detects that the set
target flow can't be reached, it gives an
alarm that the target flow can’t be
reached, and the screen will display
―Can't Reach Target Flow!!‖.
When the device works normally, if the
nasal cannula is blocked, the device will
give the nasal cannula blockage alarm,
and the screen will display ―Nasal
Cannula Blocked!!‖.