LIST OF USABLE REGISTERS
The MODBUS-RTU protocol implemented on this instrument can manage a maximum of 32
registers read and written in a single query or response.
R = the register can be read only
W = the register can be written only
R/W = the register can be both read and written
H = high half of the DOUBLE WORD forming the number
L = low half of the DOUBLE WORD forming the number
DESCRIPTION Saving to EEPROM
Divisions and Units of measure
Only after command ‘99’ of the
COMMAND REGISTER
Sample weight for calibration H
Use with command ‘101’ of the
COMMAND REGISTER
Sample weight for calibration L
40043
Weight value corresponding to
the ZERO of the analog output H
Only after command ‘99’ of the
COMMAND REGISTER
R/W
40044
Weight value corresponding to
the ZERO of the analog output L
R/W
40045
Weight value corresponding to
the Full Scale of analog output H
R/W
40046
Weight value corresponding to
the Full Scale of analog output L
R/W
CAUTION: At the time of writing, the setpoints, hysteresis values, the analog output zero and
full scale values are saved to the RAM and will be lost upon the next power-off; to store them
permanently to the EEPROM so that they are maintained at power-on, the 99 command of the
Command Register must be sent.