User manual WTS Ver.1.16
Seite 44 von 52
LIST OF USABLE REGISTERS
The MODBUS-RTU protocol implemented on this instrument can manage a maximum of 32
registers read and written in a single query or response.
R = the register can be read only
W = the register can be written only
R/W = the register can be both read and written
H = high half of the DOUBLE WORD forming the number
L = low half of the DOUBLE WORD forming the number
REGIS
TER
DESCRIPTION Saving to EEPROM
ACCES
S
40001
Firmware version
-
R
40002
Type of instrument
-
R
40003
Year of Production
-
R
40004
Serial Number
-
R
40005
Active program
-
R
40006
COMMAND REGISTER
NO
W
40007
STATUS REGISTER
-
R
40008
GROSS WEIGHT H
-
R
40009
GROSS WEIGHT L
-
R
40010
NET WEIGHT H
-
R
40011
NET WEIGHT L
-
R
40012
PEAK WEIGHT H
-
R
40013
PEAK WEIGHT L
-
R
40014
Divisions and Units of measure
-
R
40015
Coefficient H
R
40016
Coefficient L
R
40017
SETPOINT 1 H
Only after command ‘99’ of the
COMMAND REGISTER
R/W
40018
SETPOINT 1 L
R/W
40019
SETPOINT 2 H
R/W
40020
SETPOINT 2 L
R/W
40021
HYSTERESIS 1 H
40022
HYSTERESIS 1 L
40023
HYSTERESIS 2 H
R/W
40024
HYSTERESIS 2 L
R/W
40025
INPUTS
-
R
40026
OUTPUTS
NO
R/W
40037
Sample weight for calibration H
Use with command ‘101’ of the
COMMAND REGISTER
R/W
40038
Sample weight for calibration L
R/W
40043
Weight value corresponding to
the ZERO of the analog output H
Only after command ‘99’ of the
COMMAND REGISTER
R/W
40044
Weight value corresponding to
the ZERO of the analog output L
R/W
40045
Weight value corresponding to
the Full Scale of analog output H
R/W
40046
Weight value corresponding to
the Full Scale of analog output L
R/W
CAUTION: At the time of writing, the setpoints, hysteresis values, the analog output zero and
full scale values are saved to the RAM and will be lost upon the next power-off; to store them
permanently to the EEPROM so that they are maintained at power-on, the 99 command of the
Command Register must be sent.