User manual WTS Ver.1.16
Seite 45 von 52
REAL CALIBRATION COMMANDS (WITH SAMPLE WEIGHTS)
- Unload the system and reset to zero the displayed weight value with the command 100
TARE WEIGHT ZERO SETTING for calibration” of the Command Register.
- Load a sample weight on the system and send its value to the registers 40037-40038.
- To save the first sample weight value and remove the previously saved values, send the
command 101 “Save first sample weight for calibration” to the Command Register;
- To store a sample weight value and keep the previously saved values, send the command
106 “Add sample weight for calibration” to the Command Register;
- Up to 8 different sample weights can be saved to perform a linearization on multiple points.
The same sample weight can only be saved once for each calibration.
Zero sample weight values cannot be saved.
If the operation is successfully completed, the two sample weight registers are set to zero.
To cancel the real calibration and return to the theoretical calibration, send the command 104
to the Command Register. The tare reset is not cancelled.
In order to correctly set the sample weight, consider the value of the Division register
(40014). Example: to set the sample weight to 100 kg and the division is 0.001, then
the value to enter is 100000 (100 / 0.001 = 100000).