In order to correctly set a sample weight of negative value, it is necessary to consider
the registers “Sample weight for calibration” (40037 40038) as a 32-bit signed number.
If the development system does not handle signed numbers, enter the values in two’s
complement.
Example: to set the sample weight to - 56 kg, enter the values indicated in the table in
the registers “Sample weight for calibration”.
REGISTER DESCRIPTION
VALUE
HEX
DECIMAL
40037
Sample weight for instrument calibration H
0xFFFF
-1
40038
Sample weight for instrument calibration L
0xFFC8
-56
ANALOG OUTPUT SETTING
Write the weight in the registers Weight value corresponding to the Full Scale of analog output
H” (40045) and Weight value corresponding to the Full Scale of analog output L” (40046) or
write the weight in the registers weight value corresponding to the ZERO of the analog output
H” (40043) and weight value corresponding to the ZERO of the analog output L(40044). After
writing the value, send the command 99 from the Command Register to save it to EEPROM
memory.
STATUS REGISTER (40007)
Bit 0
Cell Error
Bit 1
AD Convertor Malfunction
Bit 2
Maximum weight exceeded by 9 divisions
Bit 3
Gross weight higher than 110% of full scale
Bit 4
Gross weight beyond 999999 or less than -999999
Bit 5
Net weight beyond 999999 or less than -999999
Bit 6
Bit 7
Gross weight negative sign
Bit 8
Net weight negative sign
Bit 9
Peak weight negative sign
Bit 10
Net display mode
Bit 11
Weight stability
Bit 12
Weight within +/-¼ of a division around ZERO
Bit 13
Bit 14
Bit 15