Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
97035873-01 24 of 75
5. Remove the needle stylet from the insertion needle and verify entry into the epidural space using the
standard technique.
Remove needle stylet Verify entry into the epidural space.
6. OPTIONAL. Under uoroscopic guidance, insert the lead blank through the LOR adaptor and into the
epidural space. Advance the lead blank to verify entry into the epidural space, then withdraw the blank.
7. The LOR adaptor must be removed prior to inserting the lead. Pull the LOR adaptor straight out from the
slotted needle without twisting or bending.
Recommended: Hold the slotted needle in place while removing the LOR adaptor.
Note: An anchor must be loaded over the distal end of the Innion CX Lead prior to inserting the lead
into the Entrada Needle.