ENGLISH 51 / 77 CAP. 13 - MAINTENANCE
M
ULTIGAS
CAP. 13 - MAINTENANCE
MULTIGAS 8500 PLUS is a remarkably reliable unit, manufactured using the highest quality
components, making use of the most advanced production techniques.
Please contact an authorized technical service centre for all maintenance interventions and to
purchase spare parts. The equipment includes parts subject to the PED Directive. The PED
“Pressure Equipment Directive” 97/23/EC defines and regulates all those parts that are subject to
pressure, with a specific ratio between their pressure and volume. Therefore, such parts cannot
be disassembled or handled. Under the owner’s responsibility, the equipment and its PED parts
must be verified on commissioning and regularly thereafter as provided by local legislation.
The parts subject to the PED Directive are:
RECEIVER 20 l - Category III (Dir. 97/23/EC)
SAFETY VALVE GAS AIRTEK - VS14NPT20HNBRPED4 20bar R 1/4 NPT
Category IV (Dir. 97/23/EC)
SUCTION UNIT - Art. 3.3 (Dir. 97/23/EC)
PRESSURE SWITCH 13/18bar 1/4SAE - Category IV (Dir. 97/23/EC)
The following jobs may be done by operators:
• having of the skills and qualifications provided by local legislation;
• trained in maintenance in line with the instructions given in the user and maintenance
manual.
INTERVENTIONS ON SERVICE STATION COMPONENTS WHICH ARE NOT
MENTIONED IN THE FOLLOWING PARAGRAPHS ARE PROHIBITED.
MAKE SURE THE TOOL IS UNPLUGGED FROM THE POWER MAINS BEFORE
OPENING.