CAP. 8 - COMMISSIONING 34 / 77 ENGLISH
M
ULTIGAS 8500 PLUS
CAP. 8 - COMMISSIONING
8.1 CONNECTIONS
The unit has to be positioned on a horizontal surface to ensure the correct operation.
The unit has to be connected to the electric mains following instructions on the identification plate of
the unit applied next to the main switch, mainly as to applicable voltage and power.
8.1.1 Positioning and connection
HANDLING: During handling, the minimum devices required for correct
handling shall be ensured, as provided for by accident prevention provisions.
POSITIONING: Place the unit in a stable place. The location must be well
ventilated, with a good rate of change of air. The unit must be located at least
10 cm from any potential obstacles to its internal ventilation. Keep the unit away
from rain and excessive humidity as they can irreparably damage it. In addition,
the equipment must never be directly exposed to the sunrays or to excessive
dust.
INSTALLATION: the unit must be installed by a specialised technician in
scrupulous observance of the instructions given in this manual. The use of the
equipment in explosive atmosphere is forbidden.
CONNECTIONS: since the unit is connected to the main power supply, it must
be properly grounded with its power plug GND pin. Failure to ground the unit
can damage it and constitutes a risk of fatal injury to the operator. Position the
unit so that the power plug is easy for the operator to access.