3 GENERAL
INSTRUMENTATION
3.1 Disposable Reflective Marker Spheres
General Information
Figure 1
Product Pieces Article No. Available from
Disposable Reflective Marker Spheres 90 (30 x 3) 41773 Brainlab and NDI
Disposable Reflective Marker Spheres are attached to reference arrays and instruments, thus
enabling the system to detect the position of the patient and instruments in the surgical field.
Warning
Brainlab navigation systems can only be used with the above marker spheres. The use of
other marker spheres could affect navigation accuracy, posing a risk to the patient.
When to Attach Marker Spheres
Attach marker spheres to instruments and arrays before calibration or use in surgery.
Ensuring Sterility
Disposable Reflective Marker Spheres are delivered sterile. They cannot be resterilized and
must be disposed of after use.
Warning
Do not resterilize Disposable Reflective Marker Spheres as this reduces their accuracy,
posing a risk to the patient. Disposable Reflective Marker Spheres are single use only.
GENERAL INSTRUMENTATION
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking 21