Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7 47
Software Overview
Tracking Identical
Instrument
Geometries
The software is not able to distinguish between instruments with identical tracking array ge-
ometries. For example, if you have attached a
StarLink Instrument Adapter
, size
ML
to two
different instruments (see image below), the software may interpret them as being the same
instrument.
Figure 8
Do not use pointers, microscope adapters or instrument adapters of the same
size simultaneously as the navigation software cannot distinguish between the
individual calibrations of identical geometrical arrangement of marker
spheres.