Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7 85
Surface Matching Registration
Softouch Registration
Performing Softouch Registration
Before You Begin
Carefully read the
Softouch
chapter in the
Instrument User Guide
.
Guidelines for Point
Acquisition
• Do not use the
Softouch
on soft tissue that may shift position (e.g., loose skin).
• Generally, all (bony) structures with a surface which does not generally change during
anaesthesia are appropriate for point acquisition.
• To ensure optimum accuracy, acquire points on both sides of the patient’s face.
How to Perform
Registration
Steps
1. Press
Softouch
in the
Patient Registration
dialog.
NOTE: When the
Patient Registration
dialog is open, you can also activate
Softouch
registration by holding the
Softouch
in the camera field of view.
2. Gently touch a point on the patient with the
Softouch tip and hold the instrument
still.
Acquisition of the point is indicated by a “beep” and a progress bar.
3. When a sufficient number of points is acquired,
Proceed
is enabled.
4. Press
Proceed
to start the surface matching calculation. The calculation may
take a few moments.
5. Verify registration accuracy in the dialog that opens (see page 91).