Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7 85
Surface Matching Registration
Softouch Registration
Performing Softouch Registration
Before You Begin
Carefully read the
Softouch
chapter in the
Instrument User Guide
.
Guidelines for Point
Acquisition
Do not use the
Softouch
on soft tissue that may shift position (e.g., loose skin).
Generally, all (bony) structures with a surface which does not generally change during
anaesthesia are appropriate for point acquisition.
To ensure optimum accuracy, acquire points on both sides of the patient’s face.
How to Perform
Registration
Steps
1. Press
Softouch
in the
Patient Registration
dialog.
NOTE: When the
Patient Registration
dialog is open, you can also activate
Softouch
registration by holding the
Softouch
in the camera field of view.
2. Gently touch a point on the patient with the
Softouch tip and hold the instrument
still.
Acquisition of the point is indicated by a “beep” and a progress bar.
3. When a sufficient number of points is acquired,
Proceed
is enabled.
4. Press
Proceed
to start the surface matching calculation. The calculation may
take a few moments.
5. Verify registration accuracy in the dialog that opens (see page 91).