46 Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
Tracking System
Tracking Instruments
Tracking Priority The software updates the navigation views based on the position of the instrument with the
highest priority. Lower priority instruments are displayed but are not used by default to up-
date the views.
The list shows the tracking priority, starting with high priority instruments at the top:
•
Softouch
•
Pointer
•
StarLink Instrument Adapter XL
•
StarLink Instrument Adapter L
•
StarLink Instrument Adapter ML
•
StarLink Instrument Adapter M
•
Disposable Biopsy Needle Type A
•
Biopsy Alignment Array/VarioGuide
•
Microscope Adapters
NOTE: To update the navigation views to the position of a lower priority instrument, re-
move the instruments of higher priority from the camera field of view.