Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7 63
Standard Registration
Workflow Overview
Before You Begin Fasten Disposable Registration Markers OR to the marker plates as described in the
Instrument User Guide
.
CT and MR marker sockets must be removed from the patient after the scan-
ning procedure. If markers are still attached, remove them now, taking care not
to burst the green MR markers.
Do not remove marker plates from the patient until patient registration has
been performed. Otherwise, standard registration is not possible.
Workflow
1. Select the image set (if more than one is available).
2. Plan registration points. If points were already planned in
iPlan
this step is not
necessary. Proceed to step 3.
3. Register points using either sequential or unsequential registration.
4. Verify registration accuracy.