Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7 75
Standard Registration
Next Steps Based on the precision calculated by the software and the visual verification, you can pro-
ceed as follows:
Options
If accuracy is sufficient, press
Accept
.
The main screen opens and you can begin navigation.
If a medium registration accuracy was achieved, you can:
• Press
Accept
• Press
Try Again
to repeat registration
• Press
Details
to check accuracy of individual points and/or re-acquire points that may
have been inaccurately acquired (see page 76)
NOTE: You can also check the pointer accuracy in the reference array cone as described
on page 154.
NOTE: If during visual verification a deviation of more than 3 mm is observed, we rec-
ommend that you select
Try Again
or
Details
, and/or check the pointer accuracy.
If has registration failed, press
Try Again
to repeat registration.