Output 1 and 2 Setup (OUT 1, OUT 2)
The function of open collector output 1 and 2 is configured using the Off, Pulse, Alarm, and Timer menus
detailed below.
Pulse Output (PUlse)
Pulse outputs may be assigned to rate, total, grand total, retransmit, quadrature, or test.
Rate Pulse Output (rate)
A rate based pulse output is a factor of the rate display and count (output K-factor). The rate display is a
factor of the input pulses, time base, and the input K-factor. The rate of output pulses over one time base
(seconds, minutes, hours, days) is defined below in terms of input pulses and the input K-factor and count
parameters.
For example, if the input K-factor value is set to 10, and the count set to 10, one output pulse is generated
for every 100 input pulses.
Total & Grand Total Pulse Output (Total, Grtot)
A total and grand total based pulse output is a factor of the associated total and count (output K-factor). A
pulse will be generated for every total accumulation amount equal to the count.
If the maximum output frequency would be exceeded, the meter will display the message pUlse
OVERRNG alternating on the display.