Retransmit Output (retr)
The retransmitting pulse output will send an output pulse for every input pulse, essentially duplicating the
input signal. The output will generate a pulse at the falling edge of every input pulse.
No additional programming is required for a retransmitting pulse output.
If the maximum output frequency would be exceeded, the meter will display the message pUlse
OVERRNG alternating on the display.
Quadrature Output (quad)
The pulse output set to quadrature will duplicate the other open collector output, but lag by ¼ duty cycle
(90 degrees out of phase). For example, Out 1 will follow Out 2, if Out 1 is set to quad. Only one output
should be set to quad. If both outputs are set to quad , both outputs will be disabled. The other output
should be programmed as desired for the quadrature output function, and must be a pulse (PUlse) output
selection.
Test Output (test)
The test output setting programs the output to generate pulses at a programmed constant frequency. Set
the frequency decimal point location in the DECPT menu, and then enter the desired output frequency in
Hz in the PULSE menu.
Alarm Output (alrm)
Alarm outputs may be assigned to rate, total, or grand total; or be forced on or off.