5. The meter displays inp 1. Apply a known signal and press Enter. The display will flash while accepting the signal.
6. After the signal is accepted, the meter displays dsp 1 Press Enter. Enter a corresponding display value for the signal input,
and press Enter to accept.
7. The meter displays inp 2. Apply a known signal and press Enter. The display will flash while accepting the signal.
8. After the signal is accepted, the meter displays dsp 2. Press Enter. Enter a corresponding display value for the signal input
and press Enter to accept.
9. After completing calibration the save? display will need to be acknowledged using the Enter key before calibration will take
effect.
Important Programming Note: Save (save?)
After entering the last display value, the calibration entries must be saved
(SAUE?) before they will be put into effect. However, you may move past this
selection using the Up arrow key if you need to go back and correct and earlier
entry. Once confident in the entries however, the user must navigate back to the
Save menu screen (SAUE?) and press the Enter key to save the changes.