EasyManua.ls Logo

car MLB242 - Page 23

car MLB242
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
MLB242
Users Guide MLB242-V1
Figure 15 Analysis List
Select a test you want to run by touching it on the list shown. If more test available
as shown, touch arrow-key up or down to scroll through the list.
After test selection the rst application dialogue is shown e.g. Acidity:
Cuvette with magnifying symbol
shows the reading cell to be used for
testing. It is not touchable. Position is
also indicated by a blue LED light.
Up to three samples can be tested a
run.
Figure 16 Samples for Blank Reading
Proceed as described in test application instructions, e.g. select Sample 1 and ...
A dialog appears where you can modify the sample identication, it is instructed to
pipette a reagent or sample into a cuvette. After blank reading, the cuvette icon is
marked with a green check mark. Proceed in the same way with the samples 2 and 3.
By selecting STOP, the measurement of blank values may be closed early after test
1 or 2, this happens automatically after completion of sample 3. The dialogue for the
incubation of samples is displayed then:
An instruction is indicated that should
be followed below the timer.
If you touch this question mark?”,
you will get a series of separate dia-
logs for test application instructions.
Figure 17 Sample incubation