8200-M035-O1 Rev C 200 Installation & Technical
SETUP AND CALIBRATION, CONT.
GEDGE 1650 FORMAT 4 – (Type = 10)
In setup, after selecting type 10 for the GEDGE 1650 Format 4:
At the TdAte prompt, enter the format for the date and time (0-4).
At the StrtC prompt, enter the decimal value for the start character (0-126)
At the EbLoC prompt, enter the decimal value for the end block character or data
separator (0-126).
At the dbUb prompt, enter the decimal value for the device number (0-99).
Format
<Bg><DT><BD><No><BD><W1><BD><D2><D3><D4><D5><D6><D7><BD><En>
Where:
Bg is the starting character, if 0 then it is not transmitted.
DT is the Date and Time the following BD character is output only if the
date/time is output.
BD Block division character.
D2 Displayed weight identity. G=gross, N=net
D3 M scale is in motion, S scale is stable.
D4 Inscale (0>=weight<=capacity) Overcap U=below zero
D5 Z=Center-of-Zero, otherwise ASCII space
D6 E=Tare not equal to 0, otherwise ASCII space
D7 Always an ASCII space
En End of transmission character
No Device Number 01 thru 99. If set to 0 above, then nothing is output.
W1 Displayed weight.
Weight On Demand
If continuous output has not been selected for Serial Port 1 (Cont1=NO), the 200 indicator will
respond to a weight request (ENQ).
The host device (computer) sends:
ENQ - (hex 05)
The 200 will respond:
<s><xxxxxx><d><uu><m><cc><cr>
Where:
"-" = negative, " " (blank) = positive
Added to string if enabled in setup
tn, lb, l/o, oz, t, kg, g
OC = overcap
CZ = center of zero
MO = motion
ee = weight not currently being displayed
NOTE: The Weight On Demand function is not available for Serial Port 2.