62
CALWT Readout of external weights set value for span calibration
ACALT1 Readout of ACAL time 1
ACALT2 Readout of ACAL time 2
ACALT3 Readout of ACAL time 3
P.TARE Readout of Pretare set value
ZRNG Readout of Zero Range set value
USER Readout of user unit conversion coefficient
VOL Readout of reference weight set value
DENSE Readout of surrounding liquid density set value
I.TIME Readout of Interval Timer set value
Note: The command to read out the numerical value settings can only be given by the personal
computer.
(vi) Commands for numeric value setting
CALWT= Sets external weights value for span calibration.
ACALT1= Sets ACAL time 1.
ACALT2= Sets ACAL time 2.
ACALT3= Sets ACAL time 3.
P.TARE= Sets Pretare value.
ZRNG= Sets Zero Range value.
UW= Sets unit weight.
USER= Sets user unit conversion coefficient.
VOL= Sets volume of reference weight.
SDENSE= Sets surrounding liquid density.
I.TIME= Sets interval timer value.
DATE= Sets the date.
TIME= Sets the time.
TARGET= Sets the target value.
LIMIT= Sets the limit value.
G.LO= Sets the lower limit value of Group Display mode.
G.UP= Sets the upper limit value of Group Display mode.
L.LO= Sets the lower limit value of Level Display mode.
L.UP= Sets the upper limit value of Level Display mode.
PCS= Sets the arbitrary loading piece.
#= Corresponds to numeral keys of AKB-301 Application Keyboard.
ID= Sets ID.
PASSSET= Sets PCAL password.
PASS= Inputs PCAL password.
(vii) Commands of special functions
CAL Enters Span Calibration mode.
C18 Enters Span Calibration mode.
LOCK Sets menu lock.
RELEASE Releases menu lock.
TIME Readout of date and time