36
Routine Maintenance
Routine Maintenance
CEM suggests that you perform the routine maintenance on your instrument as outlined below, along with having a
CEM Certied Technician to come check out your system once a year.
WARNING
Proper precautions must be taken to avoid contact with acids or acid vapors. Protective gear should be
worn as outlined in the user’s safety program for hazardous materials and the acid manufacturer’s safety
data sheet. Refer to these guidelines for proper handling and disposal of the acids. Dispose of all waste in
accordance with all applicable local, state, and federal health and safety recommendations.
After Every Use
• Wipe inside of MARS 6 cavity, including temperature sensors, with a lint free damp cloth.
CAUTION
Only use water and a soft cloth to clean the system. Using other cleaning solutions, brushes, and
materials can cause damage to the device.
Daily
• Ensure exhaust hose is attached
Weekly
• Clean the exhaust outlet by removing the exhaust hose and wiping the space inside the exhaust outlet with a
paper towel or disposable cloth. To clean the exhaust hose, disconnect it from the blower exhaust duct, ush it
with water and allow to dry before reconnecting to the blower duct.