Troubleshooting
Troubleshooting
Problems and Possible Solutions
Problems Possible Cause Possible Solutions
Instrument Inoperative Instrument not plugged into
electrical outlet
Plug power cord into system and a dedicated, grounded outlet.
Adequately rated power cord may be used.
Power switch not in “on” position Position power switch to on
Blown fuse Contact CEM Corporation
No Microwave Power Instrument door ajar Close MARS 6 door before pressing start
Incorrect percentage of power
selected
View “Method” section for programming methods
Low line voltage Verify line voltage is within specications
Temperature not reading
correctly
iWave sensor has debris Wipe iWave sensors with a lint free damp cloth.
Sensor needs calibration If AutoCal Calibration device is available, calibrate sensor as
outlined in this manual.
If AutoCal calibration device is not available, contact CEM
Corporation.