26. Remove the Normalization Standard, and insert the Oil
Standard into the magnet to begin Pulse Width. The
Pulse Width test will take 8-10 minutes to fully
complete. Results will be displayed in the results window
once completed.
27. The final step of System Setup is System Normalization.
The Oil standard will need to be first removed from the
ORACLE, then after approximately 20 minutes the
system will request the Normalization Standard. Place
the Normalization Standard in the magnet and press Ok.
System Normalization takes over 1 hour to complete.
Once the Normalization process is complete, the data
will be logged in the results window.
Note: Before placing the Normalization standard in the
magnet, inspect the glass swirl for any bubbles trapped
between the oil. If bubbles are present, gently shake and
rotate the tube until the bubbles are removed.