47
QuikPrep Analysis on the ORACLE with an Air or Vacuum Oven
Before beginning any sample testing, ensure the product has been fully and properly prepared in order to obtain the
best results. Improper sample preparation or tempering can reduce accuracy and repeatability. For further
information on sample preparation, please refer to the standard reference procedure, or contact CEM Corporation
Process Application Support (800) 726-3331.
QuikPrep refers to the integrated heater located on the top of the ORACLE magnet. It allows the user to rapidly
condition the sample to the necessary temperature, providing the final result much quicker than a standard
temperature conditioning with an external heater block. The QuikPrep will automatically transfer the sample into
the ORACLE for analysis at the end of its conditioning time, and is beneficial to users that require faster individual
results, but requires slightly more hands on time from the user.
When testing with an Air Oven, the user can choose to use the ORACLE to record the initial weights of the samples
that will be analyzed, or use a separate balance to record the initial weights and manually input them into the
ORACLE after the sample moisture is recorded.
1. On the ORACLE, press the menu icon and choose
Run.
2. Press the large Run arrow to begin the sample
analysis procedure