8.3.5 Setting Time Out
It sets the condition to end discharge test. Once the test time reaches the time out setting
though the measured voltage is not lower than the set conditon, the loading stops the loading
stops and so does the channel discharge test to get the capacity mAH.
Figure 8-11 Setting Time Out
8.4 Displaying Measurements
This area shows the measured voltage, current, power, execution time and the discharged
capacity at present. The voltage, current and power read back one time per second.
When the voltage is lower than Cut Off, the Channel Time Count will stop and the Capacity
(mAH) will stop counting too.
Figure 8-12 Displaying Measurements
8.5 Setting Trigger On/Off
Once all conditions are set, they can be executed by clicking Trigger On. During execution,
a Battery Voltage & Current Reading Chart will start to draw the curve according to the
measurement values to inform users the present progress. To stop it, just click Trigger Off.
All tested channels will follow the set conditions to proceed counting or stop.
Figure 8-13 Setting Trigger On/Off
8.6 SAVE AS, OPEN &BACK
The function of these three buttons is same as Hardware Setting Function; see section 4.4
and 4.5 for detail information.
8.7 Report Format
Battery test provides a simple, pure text report in *.txt format. See Section 7.10 for the
detail description of report function and its file path as well as filename.