Appendix D Troubleshooting
D-3
Simple Troubleshooting
Problem Simple Troubleshooting Procedure
Unable to turn on Power
Meter
1. Make sure the power cord is connected firmly.
2. Make sure the voltage range switch on the rear panel
matches the input power.
3. Make sure the fuse is not damaged.
Measured data error 1. Make sure the measurement wires are connected correctly.
2. Make sure the humidity condition of test environment is
within the specification.
3. Follow the steps listed in Appendix E Specification
Verification to confirm the product measurement is within
specification.
4. If the current measurement keeps showing zero after the
current is inputted, it could be the fuse on the internal current
circuit is blowout. Please contact Chroma’s dealer for
service.
Unable to do remote
control
1. Make sure the remote control address setting is correct.
2. Make sure the applicable wires are used for connection.