85
Note:
ðIt is assumed that the user has already performed the above procedure and the s c a l e
is in F PiP mode.
ðWhen the pipette calibration procedure process has not started the system will
perform normal weighing with the unit as gm.
Calibration procedure:
ðThe calibration procedure repeats itself 'n' times where 'n' is the count entered in the
count cycle settings (in FPiP settings).
ðOnce in F PiP mode, press the <CAL/MENU> key to start the calibration process.
ðOnce the <MENU/CAL> key is pressed, it will ask for the initial volume V
0.
ðFor a calibration procedure of 'n' times, the scale will ask for V 'n' times. Each of t h e s e
0
value will be stored as 'PLACE n'.
st
ðThus, for the 1 time, place the initial volume and press <CAL/MENU> key once the
stability is achieved. This is accepted as 'PLACE 1'.
ðFollow the above procedure for 'PLACE 1' to 'PLACE n'.
ðRepeat the above procedure for half volume V and full volume V each 'n' times
half full
respectively.
ðIn case of a successful calibration the scale will display 'PiP done'. If not, then it will
ask to repeat the step wrongly performed.
Volume Settings:
Initial volume V
0:
ðPress <TOGGLE> key to enter F PiP settings.
ðSelect 'V ' in F PiP settings by pressing <TOGGLE> key.
0
ðPress < TARE> to enter the 'V ' settings.
0
ðThe user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key.
ðPressing the <CAL/MENU> key will shift the cursor to the right in a cyclic way and pressing
the <TOGGLE> key will increment the digit.
ðPress the <TARE> key to select the value. This will bring you back to the F PiP menu
ðThe default volume is, V = 10.00 uL
0
Half volume V
Half :
ðPress <TOGGLE> key to enter F PiP settings.
ðSelect 'V ' in F PiP settings by pressing <TOGGLE> key.
Half
ðPress < TARE> to enter the 'V ' settings.
Half
ðThe user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key.
ð Pressing the <CAL/MENU> key will shift the cursor right in a cyclic way and pressing the
<TOGGLE> key will increment the digit.
ðPress the <TARE> key to select the value. This will bring you back to the F PiP menu
ðThe default volume is, V = 11.00 uL
Half
Full volume, V
Full. :
ðPress <TOGGLE> key to enter F PiP settings.
ðSelect 'V ' in F PiP settings by pressing <TOGGLE> key.
Full.
ðPress < TARE> to enter the 'V ' settings.
Full.
ðThe user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key.
ð Pressing the <CAL/MENU> key will shift the cursor right in a cyclic way and pressing the
<TOGGLE> key will increment the digit.
ðPress the <TARE> key to select the value. This will bring you back to the F PiP menu
ðThe default volume is, V = 12.00 UL.
Full
86