The optimum location
The correct location makes an important contribution to the accuracy of the weighing
results of high-resolution analytical and precision balances.
Hence, ensure a stable, vibration-free position as horizontal as possible.
Avoid
• Direct sunlight
• Excessive temperature fluctuations,
• Drafts (Power ----- Air Conditioning System, Fans can also cause drafts)
The best position is an a stable bench in a corner protected against drafts as far
possible from doors, windows, radiators or the ventilation slots of air conditioners.
Anti-theft device
Citizen Balance are equipped with a lug for optional anti-theft device.
The anti-theft device (cable with lock) is suitable for all models. It is available from
CITIZEN under order number CAD01.
2.1 Location
12
11
Place the components listed below inside the chamber in the order given:
(For CG Series Balance)
– Bush
– Weighing pan
Balances with a Rectangular Weighing Pan and a Weighing Capacity over 10 kg
– Place the weighing pan on the balance/scale
Under Weighing
A port for a under weighing hanger is located on the bottom of the balance.
• Remove the under weighing cover from the bottom of the balance (as shown
alongside)
• Suspend the below-balance weighing hook supplied from the hanger
• If necessary, install a shield for protection against drafts