Features
You can have the parameters pertaining to the ambient weighing conditions printed before (GLP header) and
after (GLP footer) the values of a weighing series. These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance manufacturer
– Balance model
– Balance serial number
– Software version number
– Identification number of the current sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
The record is output to a CITIZEN data printer or a computer.
Settings
Set print option to request & GLP ON
Function Keys
Press the Print key to output header and first measured value.
End an Application:
Output GLP Footer : Press Cancel Key
End an application program Press Cancel key
------------------------ Dotted line
28-Jul-10 03:19PM Date / Time (beginning of measurement)
Citizen Balance Manufacturer
Model CX 220 Balance Model
Ser.no. 9223102 Balance Serial Number
Ver.no. r0.1.5.3 Software Version
ID 1234567 ID
------------------------ Dotted line
LID: 1111111 Lot ID
nRef 170 pcs Counting : Reference Sample Quantity
wRef 0.2945 g Counting : Reference Weight
Qnt + 170 pcs Counting Result
+ 50.0650 g Weighing Result
+ 250.3250 ct Weighing Result
------------------------ Dotted line
28-Jul-10 03:23PM Date / Time (end of measurement)
Name: Name of Operator
................
------------------------ Dotted line
------------------------ Dotted line
20-Jul-10 10:32AM Date / Time (beginning of measurement)
Citizen Balance Manufacturer
Model CX 220 Balance Model
Ser.no. 9223102 Balance Serial Number
Ver.no. r0.1.5.3 Software Version
ID 1234567 ID
------------------------ Dotted line
Calibration: External Calibration / Adjustment Mode
Blank Line
W-ID
Temperature
Set + 200.0000 g Calibration Weight
Diff. + 0.1234 g Diff. After Calibration
External Cal Done Confirmation of Completed Calibration
Blank Line
Diff. 0.0000g Difference from Nominal Value after Calibration
------------------------ Dotted line
20-Jul-10 10:32AM Date / Time (end of measurement)
Name: Name of Operator
................
------------------------ Dotted line
The ISO/GLP-compliant record can contain the following lines:
............ Weight ID
32.905'C Temperature
The ISO/GLP-compliant record can contain the following lines:
105
106
8. ISO/GLP-compliant Printout/Record